Description

The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.

References

https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01

https://www.cisa.gov/news-events/alerts/2025/01/30/cisa-releases-fact-sheet-detailing-embedded-backdoor-function-contec-cms8000-firmware

https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication

https://www.cisa.gov/resources-tools/resources/contec-cms8000-contains-backdoor

https://www.bleepingcomputer.com/news/security/backdoor-found-in-two-healthcare-patient-monitors-linked-to-ip-in-china/

Details

Source: Mitre, NVD

Published: 2025-01-30

Updated: 2025-03-01

Risk Information

CVSS v2

Base Score: 7.6

Vector: CVSS2#AV:N/AC:H/Au:N/C:C/I:C/A:C

Severity: High

CVSS v3

Base Score: 7.5

Vector: CVSS:3.0/AV:N/AC:H/PR:N/UI:R/S:U/C:H/I:H/A:H

Severity: High

CVSS v4

Base Score: 7.7

Vector: CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N

Severity: High